844-36 Process Monitor Calibration Fixture
- Reproducible calibration geometry
- 2 inches of 4 Pi lead shielding to reduce background response
- Reversible source drawer for Beta and Gamma scintillation detector calibration
- Scintillation detector dead time verification drawer available
- GM detector adaptors available
- Disassembles easily for field transportation
The gain and low discriminator settings directly affect the efficiency of the detector. Because the gain of the PMT used in scintillation detectors decreases with age, it is necessary to periodically readjust the detector high voltage to ensure the gain the detector is operating at is the same as the gain the detector was operating at when the detector’s primary calibration was performed. The frequency of re-calibration is a function of the individual application, and should be performed at 12 to 24 month intervals. For process Geiger-Mueller (GM) detectors, a similar, although not as critical (due the wide high voltage operating region) re-calibration process is also required. To permit field re-calibration of scintillation and process GM detectors, Victoreen has designed the Model 844-36 Process Monitor Calibration Fixture. The Model 844-36 is also referred to as a Standard Geometry. The Model 844-36, and its associated reference point sources, may be used as a secondary standard, to perform NIST traceable field re-calibrations of process monitor scintillation and GM detectors.
- Lead Shielded Calibration Fixture
- Field Calibration of Scintillation and Process GM Detectors
- Process Monitor Detector Calibration
- Permits NIST Traceable Detector Re-calibrations in the Field
The Model 844-36 and associated reference sources are used to field calibrate Fluke Biomedical’s pulse mode process monitor scintillation and GM detectors. The Model 844-36 consists of three (3) lead filled “doughnuts” clamped together with four (4) bolts, to provide a lead shielded cylinder, into which a scintillation detector or a GM detector (when installed in the appropriate adaptor) may be inserted. The assembled calibration fixture provides two (2) inches of lead shielding to minimize the detector response induced by ambient background radiation. At the bottom of the fixture, a removable slide drawer is provided to permit installation and withdrawal of reference calibration disc sources. The drawer is reversible, with one side machined for beta sources and the other side machined for gamma sources. An optional tau drawer, machined to accept the two sources required to measure detector dead time, is also available. A stop is provided in the fixture to position the detector at a fixed, reproducible distance from the reference disc source. To perform a field re-calibration, the detector is inserted into the Model 844-36, the appropriate calibration disc source is inserted into the slide drawer, and the counts from the detector for a fixed time period are recorded. To achieve a 1% counting error, the counting time should be long enough to yield a minimum of 10,000 counts. A number if field calibration sources are available for field calibrating both beta and gamma scintillation detectors. For GM detectors, the detector is inserted into a Lucite® adaptor designed for the detector, the reference source is inserted into a cavity in the adaptor, and the adaptor, with GM detector and reference source is inserted into the Model 844-36. In this case, the Model 844-36 functions as a shield, to reduce ambient background radiation.
Model Name |
Description |
844-36 | Process Monitor Calibration Fixture |
Miscellaneous Accessories |
|
Beta disc sources |
Beta disc sources (part# 844-13-11)14C, 0.1 \u00b5Ci (approximately) Model number: 844-13-11 |
Tau disc source |
Tau disc source (part# 844-13-16)90Sr (set of 2), 20.0 \u00b5Ci (approximately)*Requires copy of User's Material License prior to shipment Part number: 844-13-16 Model number: 844-13-16 |
Gamma Disc Source |
Gamma Disc Source (part# 844-13-6)137Cs, 0.5 \u00b5Ci (approximately) Model number: 844-13-6 |
Gamma Disc Source |
Gamma Disc Source (part# 844-13-7)133Ba, 0.5 \u00b5Ci (approximately) Model number: 844-13-7 |
844-13-9 Beta disc sources |
844-13-9 Beta disc sources (part# 844-13-9)90Sr, 0.1 \u00b5Ci (approximately) Model number: 844-13-9 |
844-36-15 Gamma Disc Source |
Gamma Disc Source (part# 844-36-15)133Ba (set of 3), 0.015, 0.15, 1.5 \u00b5Ci (approximately) Model number: 844-36-15 |
844-36-16 Gamma Disc Source |
Gamma Disc Source (part# 844-36-16137Cs (set of 3), 0.004, 0.04, 4.0 \u00b5Ci (approximately) Model number: 844-36-16 |
844-36T Calibration Fixture |
844-36T Calibration Fixture (part# 844-36T) includes:Tau source holderTau source spacerSpacer removal rod Model number: 844-36T |
844-36-17 Beta disc sources |
Beta disc sources (part#844-36-17)137Cs (set of 3), 0.001, 0.01, 0.1 \u00b5Ci (approximately) |